TopPette
轻便且设计符合人机工效学;
数字视窗,令所设定量程一目了然;
量程范围广,0.1μL至10mL;
使用附件工具,能方便快捷地进行校准和维修;
精确的分液,每支移液器都遵照EN/ISO8655标准进行校准;
下半支可高温高压消毒
技术参数
单道可调式移液器 |
||||||
量程 |
增量 |
测量体积(μl) |
允许最大系统误差 |
允许最大随机误差 |
||
|
|
|
% |
μl |
% |
μl |
0.1-2.5μl |
0.05μl |
2.5 |
2.50% |
0.0625 |
2.00% |
0.05 |
1.25 |
3.00% |
0.0375 |
3.00% |
0.0375 |
||
0.25 |
12.00% |
0.03 |
6.00% |
0.015 |
||
0.5-10μl |
0.1μl |
10 |
1.00% |
0.1 |
0.80% |
0.08 |
5 |
1.50% |
0.075 |
1.50% |
0.075 |
||
1 |
2.50% |
0.025 |
1.50% |
0.015 |
||
2-20μl |
0.5μl |
20 |
0.90% |
0.18 |
0.40% |
0.08 |
10 |
1.20% |
0.12 |
1.00% |
0.1 |
||
2 |
3.00% |
0.06 |
2.00% |
0.04 |
||
5-50μl |
0.5μl |
50 |
0.60% |
0.3 |
0.30% |
0.15 |
25 |
0.90% |
0.225 |
0.60% |
0.15 |
||
5 |
2.00% |
0.1 |
2.00% |
0.1 |
||
10-100μl |
1μl |
100 |
0.80% |
0.8 |
0.15% |
0.15 |
50 |
1.00% |
0.5 |
0.40% |
0.2 |
||
10 |
3.00% |
0.3 |
1.50% |
0.15 |
||
20-200μl |
1μl |
200 |
0.60% |
1.2 |
0.15% |
0.3 |
100 |
0.80% |
0.8 |
0.30% |
0.3 |
||
20 |
3.00% |
0.6 |
1.00% |
0.2 |
||
50-200μl |
1μl |
200 |
0.60% |
1.2 |
0.15% |
0.3 |
100 |
0.80% |
0.8 |
0.30% |
0.3 |
||
50 |
1.00% |
0.5 |
0.40% |
0.2 |
||
100-1000μl |
5μl |
1000 |
0.60% |
6 |
0.20% |
2 |
500 |
0.70% |
3.5 |
0.25% |
1.25 |
||
100 |
2.00% |
2 |
0.70% |
0.7 |
||
200-1000μl |
5μl |
1000 |
0.60% |
6 |
0.20% |
2 |
500 |
0.70% |
3.5 |
0.25% |
1.25 |
||
200 |
0.90% |
1.8 |
0.30% |
0.6 |
||
1000-5000μl |
50μl |
5000 |
0.50% |
25 |
0.15% |
7.5 |
2500 |
0.60% |
15 |
0.30% |
7.5 |
||
1000 |
0.70% |
7 |
0.30% |
3 |
||
2-10ml |
0.1ml |
10ml |
0.60% |
60 |
0.20% |
20 |
5ml |
1.20% |
60 |
0.30% |
15 |
||
2ml |
3.00% |
60 |
0.60% |
12 |